THE EFFECTS OF CHANGING MEMBRANE COMPOSITIONS AND INTERNAL ELECTROLYTES ON THE RESPON OF POTASSIUM ION SENSOR
Abstract
A study on the changing of membrane compositions and internal solution towards the response potassium ion sensor was carried out. Potassium ion sensor based on photocured cross linking poly(n-butyl acrylate) membranes with varying composition of valinomycin (val), sodium tetrakis [3.5-bis(trifluoro-methyl) phenyl] borat (NaTFPB), types ion of internal solution were investigated. Effects of varying composition of val, NaTFPB, types and concentration of internal solution were observed on potassium ion sensor. The effect of higher val composition was lower LOD, wider linear range, lower sensitivity but increase selectivity. Higher NaTFPB compositions lead to lower LOD, higher sensitivity and selectivity. The effect of changing internal electrolyte lead to lower LOD, wider linear response range and higher selectivity according to internal electrolyte consisting Na+, Ca2+ and K+, and sensitivity increase following internal electrolyte with Ca2+, K+ and Na+.
Key words: ion sensor, ion selective membrane, ionofor, lipofilik saltFull Text:
PDF (Bahasa Indonesia)References
Abranova, N.Y. Bratov, A.V. Alvasov, Y.G. & Bartroli, D. 1998. Photopolymerized polyacrylates as Ion Selective Lectrodes of Chemical Sensor. J. Anal. Chem. 53(8): 756-761.
Anon. Valinomycin. Darai halaman website: http://en.wikipwedi.org/wiki/valinomycin [28 Maret 2007].
Eric, B. & Pretsch, E. 2001. Potentiometriy at trace level. T. Anal. Chem. Vol. 20. No.1.
Eric, B. & Pretsch, E. 2005. Potentiometric sensors for trace-level analysis. T. Anal. Chem. Vol.24,No.3.
Joseph W. 2001. Analytical electrochemistry. New York: John Wiley & Son.
Lee. Y.H. & Hall, E.A.H. 1996. Methacrylate-Acrylate Based Polymer of Low Plasticiser Content for Potassium Ion Selective Membranes. Anal. Che. Acta 324: 47-56.
Lee. Y.H. & Hall, E.A.H. 2000. Assesing a Photocured Self-Plasticised Acrylic Membrane Recipe for Na+ and K+ Ion Selective Electrodes. Anal. Chem. 443: 25-40.
Lee. Y.H. & Hall, E.A.H. 2001. Methacrylic-Acrylic Polymer in Ion-Selective Membranes: Achieving the Right Polymer Recipe. Anal. Chem. 403: 77-89.
Lee. Y.H. 1998. Ion-selective acrylic polymer matrices. London: Magdalena College. Institute of Biotechnology. University of Cambridge.
Lee. Y.H., Musa A, Nasir M.A. & Sagir. A. 2004 Pengembangan sensor kalium dengan teknik fotopolimer dan elektroda bercetak skrin. Yokyakarta. Seminar Nasional Kimia XIV.
Moerf, W.E. & Simon W. 1978. Ion-Selective Electode Based on Neutral Carrier in Ion-Selective Electrode in Analytical Chemistry. Vol I, Freiser H., New York: Plenum Press.
Wojciech, W. 2005. Ion-selective electrod. Atas talian:www.csrg.ch.pw.edu.pl [21 Pebruari 2007].
DOI: http://dx.doi.org/10.31958/js.v7i1.119
Refbacks
- There are currently no refbacks.
Copyright (c) 2016 Alizar Ulianas, Lee Yook Heng
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Indexed by:
__________________________________________________________________________
Sainstek: Jurnal Sains dan Teknologi
ISSN 2085-8019 (print) | 2580-278x (online)
Published by Institut Agama Islam Negeri Batusangkar
Email: sainstek@iainbatusangkar.ac.id
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.